Agilent Technologies 8590 Series Service Manual page 624

Analyzers assembly-level repair
Table of Contents

Advertisement

Analyzer Messages
Interpreting Analyzer Messages
CAL: MAIN COIL SENSE FAIL (H)
The analyzer could not set up span sensitivity of the main coil.
Related Assemblies
8590L and 8591E spectrum analyzers only: A6 YTO, A7 analog
interface.
All other 8590 E-Series and L-Series spectrum analyzers: A3A7 YTO,
A7 analog interface.
Troubleshooting Hints
If this message appears, press
,
More 2 of 4
DEFAULT CAL DATA
again.
CAL: NBW 200 Hz notch amp failed (H)
Indicates that the 200 Hz resolution bandwidth is not the correct shape
for the calibration routine.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the
CAL AMPTD
CAL: NBW 200 Hz notch failed (H)
Indicates that the 200 Hz resolution bandwidth is not the correct shape
for the calibration routine.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the
CAL AMPTD
624
, −37,
FREQUENCY
, and perform the
.
.
,
,
,
Hz
CAL
More 1 of 4
routine
CAL FREQ
Chapter 14