Agilent Technologies 8590 Series Service Manual page 140

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments
22. First LO Distribution Amplifier for Option 009 or 010
Figure 2-33
First LO Distribution Amplifier Adjustment Setup
4. Set the digital voltmeter controls as follows:
5. Set the analyzer
6. Adjust A10R29 (GATE) for a digital voltmeter reading within 5 mV
of the GATE (gate bias) voltage printed on the RF section label.
7. Zero and calibrate the measuring receiver and power sensor in LOG
mode. (Power levels read in dBm.) Enter the power sensor's 5 GHz
cal factor into the measuring receiver.
8. Connect the power sensor to the analyzer LO OUTPUT.
9. On the analyzer, press
MHz
10.Connect the positive DMM test lead to A10TP4, LOS (LO sense).
11.Note the SENS (LO sense) voltage printed on the RF section label.
Adjust A10R25, LO AMP (LO power), until the DMM reads equal to
the SENS voltage printed on the RF section label.
140
FUNCTION .............................................. DC VOLTS
RANGE ................................................................ 10 V
RESOLUTION .................................................... 1 mV
switch to on.
LINE
PRESET
.
,
,
,
SPAN
ZERO SPAN
FREQUENCY
, 300,