Agilent Technologies 8590 Series Service Manual page 158

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments
26. Correcting for Absolute Amplitude Accuracy for Option 050
4. Read the amplitude error from the display and record it in
Table 2-15
Frequency Response Errors: 8593E, 8594E, 8595E, 8596E
Frequency
804 MHz
876 MHz
948 MHz
1.02 GHz
Table 2-16
Frequency Response Errors: 8591C and 8591E
Frequency
781 MHz
818 MHz
855 MHz
892 MHz
5. Subtract the Average Error, calculated in
6. Using
158
2-15
for the 8593E, 8594E, 8595E, and 8596E spectrum analyzers.
Use
Table 2-16
for 8591E spectrum analyzers or 8591C cable TV
analyzers.
Amplitude ERROR
Amplitude ERROR
Repeat
step 3
and
step 4
Table 2-15
or
Table 2-16
Amplitude ERROR from
column 3 as the Corrected Flatness ERROR.
Do not subtract more than 0.5 dB from the Amplitude ERROR. If the
Average Error is more than 0.5 dB, it may not be possible to make
corrections during the adjustment.
(step down key), set the frequency displayed on the analyzer
to 804 MHz (781 MHz for 8591E spectrum analyzers or 8591C cable
TV analyzers).
Corrected Flatness
Corrected Flatness
for the next frequency indicated in
until all four points have been recorded.
step
Table 2-15
or
Table
2-16, then record
Table
ERROR
ERROR
1, from each
Chapter 2