Agilent Technologies 8590 Series Service Manual page 619

Analyzers assembly-level repair
Table of Contents

Advertisement

Table 14-2
Possible Cause of the Error Message ADC-2V FAIL
A16J1
Pin Number
1
3
5
7
31
Figure 14-2
A16J1 Connector-Pin Orientation
Chapter 14
Possible Error-Message Source
A14 log amplifier assembly An assembly installed in the card cage
An assembly installed in the card cage
An assembly installed in the card cage
A16 processor/video assembly
An assembly installed in the card cage
Signal incorrectly applied to the AUX VIDEO output
A7 analog interface assembly
Analyzer Messages
Interpreting Analyzer Messages