4. On the analyzer, press the following keys.
PRESET
CAL, More 1 of 4
CORRECT ON OFF
, 300,
FREQUENCY
CF STEP AUTO MAN
, 12,
SPAN
MHz
75 Ω input: press
, 10,
AMPLITUDE
SCALE LOG LIN
, 1,
BW
MHz
PEAK SEARCH
,
MKR FCTN
MK TRACK ON OFF
5. Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK
amplitude reading of −14 dBm ±0.1 dB.
6. Press RATIO mode on the measuring receiver.
7. Press the following analyzer keys.
, 41,
FREQUENCY
8. Set the synthesized sweeper CW to 41 MHz.
9. Adjust the synthesized sweeper POWER LEVEL for an analyzer
MKR-TRK amplitude reading of −14 dBm ±0.1 dB.
10.Record the power ratio here and in Column 2 of
41 MHz.
Measuring Receiver Reading at 41 MHz____________________dB
11.Press the following analyzer keys.
, 78,
FREQUENCY
12.Set the synthesized sweeper CW to 78 MHz.
13.Adjust the synthesized sweeper POWER LEVEL for an analyzer
MKR-TRK amplitude reading of −14 dBm ±0.1 dB.
14.Record the power ratio displayed on the measuring receiver in
Column 2 of
Table 2-6
Chapter 2
15. Frequency Response of the 8590L, 8591C, and 8591E
(OFF)
MHz
, 37,
MHz
,
AMPLITUDE
More 1 of 2
−dBm
(LOG), 1,
dB
(ON)
MHz
MHz
for 78 MHz.
Making Adjustments
,
,
AMPTD UNITS
dBm
Table 2-6
for
99