Making Adjustments
15. Frequency Response of the 8590L, 8591C, and 8591E
15.On the analyzer, press
16.On the synthesized sweeper, press CW and STEP UP.
Frequency Response Error At 4 MHz
17.Using a cable, connect the frequency synthesizer directly to the
Figure 2-29
Frequency Response for 4 MHz Setup
Use only 75 Ω cables, connectors, or adapters on instruments equipped
CAUTION
with 75 Ω inputs or damage to the input connectors will occur.
100
the remaining frequencies listed in Column 1 of
new frequency repeat
Cal Factor into the measuring receiver as indicated in Column 3 of
Table
2-6.
INPUT 50Ω. See
Figure
connect the frequency synthesizer 75Ω OUTPUT to the INPUT 75Ω
of the analyzer. Set the frequency synthesizer 50–75 Ω switch to the
75 Ω position. See
Figure
and
(step up), to step through
FREQUENCY
step 13
to
step
15, entering the power sensor
2-29. For 75 Ω inputs use a 75 Ω cable to
2-29.
Table
2-6. At each
Chapter 2