Agilent Technologies 8590 Series Service Manual page 193

Analyzers assembly-level repair
Table of Contents

Advertisement

6. Record the power meter reading in Column 4 of
7. On the synthesized signal generator, press CW and STEP UP, to step
Procedure
1. Zero and calibrate the measuring receiver and 1 MHz to 1.8 GHz
2. Connect the equipment as shown in
Figure 2a-5
Frequency Response Setup
Use only 75 Ω cables, connectors, or adapters on instruments equipped
CAUTION
with 75 Ω inputs or damage to the input connectors will occur.
3. Press INSTRUMENT PRESET on the synthesized signal generator.
Chapter
into account the cal factors of both power sensors.
through the remaining frequencies listed in
At each new frequency repeat
power sensor cal factor into the respective power meter.
power sensor in log mode as described in the measuring receiver
operation manual.
Set the synthesized signal generator controls as follows:
CW ................................................................ 300 MHz
FREQ STEP ................................................... 37 MHz
POWER LEVEL ............................................. −9 dBm
2a
Making Adjustments: If 3335A Source Not Available
15a. Frequency Response of the 8590L, 8591C, and 8591E
Table
step 5
and
step
Figure
2a-5.
Table
2a-5, taking
2a-5.
6, and enter each
193