Troubleshooting the RF Section
6b. 8592L/94L, 8594Q, and 8593E/94E/95E/96E
Table 6-4
Power Levels at Measurement Points
Measurement
Measurement
Point
Frequency
A
Same as input
frequency
B
Same as input
frequency
C
321.4 MHz
D
3.9214 GHz
E
321.4 MHz
* A frequency-selective measuring device, such as another analyzer, is recommended for making these
measurements. Broadband measuring devices, such as power meters, will give erroneous results due to the
presence of other, higher-amplitude signals.
† The power-level range is relative to the actual measurement taken at the measurement point indicated.
288
Band 0
Band 1
92L, 93E,
92L, 93E,
94E, 94L,
95E, 96E
94Q, 95E,
96E
−30 to −33
dBm
−14 to −19
dBm
10 to 16 dB
below B †
8 to 12 dB
below A †
5 to 11 dB
1 to 6 dB
below D †
below C †
Power Level Range *
Band 2
Band 3
92L, 93E,
92L, 93E
96E
−14 to −19
−14 to −20
dBm
dBm
18 to 26 dB
21 to 31 dB
below B †
below B †
1 to 6 dB
1 to 6 dB
below C †
below C †
Band 4
92L,93E
−15 to −23
dBm
25 to 35 dB
below B †
1 to 6 dB
below C †
Chapter 6