Making Adjustments
16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
25.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK
amplitude reading of −9 dBm ±0.1 dB on the analyzer.
26.Record the power ratio displayed on the measuring receiver in
Column 2 of
Table
27.On the synthesized sweeper, press CW and STEP UP.
28.On the analyzer, press the following keys.
FREQUENCY, ⇑
29.Step through the remaining frequencies listed in
new frequency, repeat
power sensor cal factor into the measuring receiver as listed in
Column 3 of
Table
is complete. Continue with "Entering Flatness Correction Data" at
the end of the adjustment.
Continue with the next step for all other analyzers.
Band 2
30.To enter band 2, press the following analyzer keys.
,
FREQUENCY
BAND LOCK
31.Set the analyzer controls by pressing the following keys.
, 6.2,
FREQUENCY
CF STEP AUTO MAN
, 10,
SPAN
MHz
, 1,
BW
MHz
32.Set the synthesized sweeper CW to 6.2 GHz and FREQ STEP to
184 MHz.
33.Press the following analyzer keys.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF
AMPLITUDE, PRESEL PEAK
34.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK
amplitude reading of −15 dBm ±0.1 dB on the analyzer.
35.Add 6 dB to the power ratio displayed on the measuring receiver and
record the result in
For example, if the power ratio displayed on the measuring receiver
is +2.8 dB, enter 8.8 dB (Error = 2.8 dB + 6.0 dB = 8.8 dB). The 6 dB
correction comes from the difference between the MKR-TRK
amplitudes set in
110
2-9.
step 24
to
step 28
2-9.Adjustment for the 8595E spectrum analyzer
,
6.0-12.8 BAND 2
GHz
(MAN), 184,
MHz
(ON)
Table
2-11.
step 6
(−9 dBm) and
Table
2-9. At each
and enter the appropriate
step 34
(−15 dBm).