CAL: NBW 200 Hz width failed (H)
Indicates that the 200 Hz resolution bandwidth is not the correct
bandwidth for the calibration routine.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the
CAL AMPTD
CAL: NBW gain failed (H)
Indicates that one of the resolution bandwidths is not the correct
amplitude for the calibration routine.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the
CAL AMPTD
CAL: NBW width failed (H)
Indicates that one of the resolution bandwidths is not the correct width
for the calibration routine.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the
CAL AMPTD
CAL: PASSCODE NEEDED (M)
The service function executed cannot be performed without the service
passcode.
Troubleshooting Hints
Before performing any function that requires a service passcode, be
sure you understand the consequences of the function. Improper use of
functions that require a passcode can delete factory correction
constants that are difficult to replace.
Refer to the passcode information in
the proper use of the passcodes.
Chapter 14
.
.
.
Chapter 13
Analyzer Messages
Interpreting Analyzer Messages
for information about
625