Making Adjustments
21. Correcting for External ALC Error for Option 010 or 011
Procedure
1. Connect a BNC cable from the RF INPUT to the RF OUTPUT.
Use only 75 Ω cables, connectors, or adapters on instruments equipped
CAUTION
with 75 Ω inputs or damage to the input connectors will occur.
2. Press
3. Perform the tracking generator self-cal routine by pressing the
4. Press
5. Zero and calibrate the power meter and power sensor as described in
Use only 75 Ω cables, connectors, or adapters on instruments equipped
CAUTION
with 75 Ω inputs or damage to the input connectors will occur.
Figure 2-32
External ALC Error Correction Adjustment
136
, then wait for preset routine to finish.
PRESET
following analyzer keys.
,
CAL, More 1 of 4
More 2 of 4
Wait for the completion of the routine (less than 1 minute).
.
CAL STORE
the power meter operation manual.
,
CAL TRK GEN
Chapter 2