Agilent Technologies 8590 Series Service Manual page 634

Analyzers assembly-level repair
Table of Contents

Advertisement

Analyzer Messages
Interpreting Analyzer Messages
The 10-Digit Failure Code
The 10-digit segment provides further failure information for the I/O
address bus, the I/O data bus, and the A16 video RAM circuitry. Each
digit of the 10-digit hexadecimal segment must be translated to its
binary equivalent to obtain the failure information.
The 10-digit failure code section is divided into three separate codes.
The 10-digit failure code example (0625C1F72D) is separated into three
codes as shown below.
FAIL: DF40 (06) (25C1) (F72D)
Each failure code from the example is converted from hexadecimal to
binary and interpreted in
The I/O Address Bus Failure Code.
The first two digits of the 10-digit code identify the failed address lines
on the I/O address bus. Convert the two digits to binary and interpret
them with
Table 14-4
I/O Bus Address Failure-Code Interpretation
Digits From
Example
634
Table
Table
14-4.
Equivalent
Binary Value
8
4
2
0→
1
8
6→
4
2
1
14-4,
Table
14-5, and
A16 Circuit Tested
Not Used
Not Used
Not Used
Error at ADR 4
Error at ADR 3
Error at ADR 2
Error at ADR 1
Error at ADR 0
Table
14-6.
A16 Reference
Designator
U18
U18
U18
U18
U18