Agilent Technologies 8590 Series Service Manual page 643

Analyzers assembly-level repair
Table of Contents

Advertisement

Analyzer Messages
Interpreting Analyzer Messages
INVALID REPEAT NEST LEVEL (U)
The nesting level in the REPEAT routine is improperly constructed.
This can occur if too many REPEAT routines are nested. When used
within a downloadable program (DLP), the maximum number of
REPEAT UNTIL statements that can be nested is 20.
INVALID RS-232 ADRS/OPERATION (U)
An RS-232 operation was aborted due to an invalid operation.
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
Reset all instruments attached to the bus.
INVALID SAVE REG (U)
Data has not been saved in the specified state or trace register, or the
data is corrupt.
INVALID SCRMOVE (H)
Indicates the analyzer may have a hardware failure.
Related Assemblies
All internal assemblies.
Troubleshooting Hints
Remove instrument from bus. Perform an instrument preset.
INVALID START INDEX (U)
Indicates that the first trace element specified for a range of trace
elements is not within the trace range of the specified trace.
INVALID STOP INDEX (U)
Indicates that the ending trace element specified for a range of trace
elements is not within the trace range of the specified trace.
INVALID STORE DEST: _ _ _ (U)
The specified destination field is invalid.
Chapter 14
643