Agilent Technologies 8590 Series Service Manual page 582

Analyzers assembly-level repair
Table of Contents

Advertisement

Softkey Descriptions
Softkey Descriptions
DEFAULT CAL DATA Default Calibration Data
D
The current correction-factor data is replaced in
ESCRIPTION
nonvolatile memory by the factory-loaded default
calibration data when
default data can be used only if a passcode, either
−37 Hz or −2001 Hz, has been entered.
There are two types of correction factors that are
changed when the default data is used.
• Video offsets. They are used on the A16
• DAC settings. They are used on A7 analog interface
After entering the passcode and pressing
DATA
CAL DATA
the screen display of the default-calibration data.
Figure 13-5
analyzers equipped with Option 130.
Note that the default calibration data provided below is
for an 8593E. The default calibration data for your
analyzer may be slightly different.
Figure 13-3 Default Calibration Data: Page 1
582
DEFAULT CAL DATA
processor/video assembly.
assembly.
, the default data can be viewed using
.
Figure 13-3
through
provides additional information for
is used. The
DEFAULT CAL
DISPLAY
Figure 13-4
provides
Chapter 13