Earth Leakage Current; Enclosure Leakage Current; Patient Applied Risk Current - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Specifications

Earth Leakage Current

Enclosure Leakage Current

Patient Applied Risk Current

102
AC Polarity
Line Cord
Normal
Closed
Reversed
Closed
Normal
Open
Normal
Closed
AC Line
Neutral
Polarity
Line Cord
Normal
Closed
Normal
Closed
Normal
Open
Reversed
Closed
Reversed
Open
Reversed
Closed
* AAMI/ANSI-ES1 does not include opening the line conductor.
AC Line
Neutral
Polarity
Line
Normal
Closed
Normal
Open
Normal
Closed
Reversed
Closed
Reversed
Open
IEC 60601-1
Neutral
AAMI/
Cord
ANSI-ES1
Closed
500 µA
Closed
500 µA
Closed
1000 µA
1000 µA
Open
Power
Line
IEC
Ground
60601-1
Cable
100 µA
Closed
Open
500 µA
Closed
100 µA
Closed
500 µA
500 µA
Closed
Open
500 µA
Power Line
IEC
Ground
60601-1
Cable
UL 2601-1
Closed
100 µA
Closed
500 µA
Open
500 µA
Closed
100 µA
500 µA
Closed
UL 2601-1
300 µA
300 µA
500 µA
500 µA
AAMI/ANSI
-ES1 UL 2601-1
300 µA
300 µA
300 µA *
300 µA
300 µA *
300 µA
AAMI/ANSI-ES1
10 µA **
50 µA **
50 µA **
10 µA **
50 µA **