Table 14: Cable Compliance - Nellcor OxiMax N-550B Service Manual

Table of Contents

Advertisement

Specifications
Manufacturer's Declaration
86
Item
Safety Alarms
EN 60601-1-8: 2004
Front panel and
IEC 60878, EN 980,
case labeling
ISO 7000, EN 60417-1,
EN 60417-2
Button spacing
ISO 7250
Year of
EN 980
manufacture
symbol
Operation during
IEC 60068-2-27
physical shock
Operation during
IEC 60068-2-6, IEC
vibration
60068-2-34
This product complies with EN/IEC60601-1-2:2001.
WARNING: The use of accessories, sensors, and cables other than those
specified may result in increased emission and/or create invalid readings
of the N-550B.

Table 14: Cable Compliance

RF emissions, EN 55011, Class B/Group 1
EN 60601-1-2; 2001
Cables and OxiMax Sensors
OxiCliq-OC-3 cable
DOC-10 pulse oximeter cable
Software download cable, RS-232 serial, 15 to 9 pin "D"
Non-terminated cable, RS-232/Analog, 15 pin "D"
Oxinet hardwire cable
Printer cable, RS-232, 15 to 9 pin "D"
Standard
General Application of Alarms in
Medical Electric Equipment
Cables Comply With:
Description
Maximum Length
3 ft. (0.9 m)
10 ft. 10 in. (3.3 m)
10 ft. (3 m)
3.3 ft. (1 m)
10 ft. (3 m)
10 ft. (3 m)

Advertisement

Table of Contents
loading

Table of Contents