Replacing The Battery - Nellcor OxiMax N-600x Service Manual

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Replacing the Battery

O
M
N-600x Service Manual
XI
AX
5. Be sure to note the polarity of the leads. Use needle-nose pliers to disconnect
the leads from the battery.
Figure 7: Removing the Battery
Caution: The lead-acid battery is recyclable. Do not dispose of the battery by
placing it in the regular trash. Dispose of the battery in accordance with local
guidelines or contact Nellcor's Technical Services department to arrange for
disposal.
To replace the old battery:
1. Follow the steps outlined in Removing the Old Battery on page 88.
2. Connect the leads to the battery. The red wire connects to the positive
terminal, and the black wire connects to the negative terminal.
3. Insert the new battery into the bottom case with the negative terminal towards
the outside of the monitor. Install the bracket and grounding lead with the two
screws.
Disassembling the Monitor
89