Electrical; Environmental Conditions - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Specifications

Electrical

Conditions
Environmental
90
Instrument
Power Requirements
Fuses
Battery
The battery provides at least 2 hours of battery life when new and fully charged with no
alarms, no serial data, no analog output, no nurse call output, with backlight on while
using a pulse simulator set for 224 bpm, high light and low modulation.
Type
Voltage
Recharge
Shelf Life
Complies With
O
M
Sensors
XI
AX
Wavelength and
The wavelength range of the light emitted are near 660 nm and
Power
890 nm with the energy not exceeding 15 mW.
Operating
Temperature
Altitude
Atmospheric Pressure
rated at 108 to 132 volts AC (nominal 120 VAC) or 200
to 240 volts AC (nominal 230 VAC), 20 volt/amps to be
compliant with IEC 60601-1 sub-clause 10.2.2
qty 2, 0.5 A, 250 volts, slow-blow, IEC (5 x 20 mm)
Lead acid
6 Volts DC
14 hours with N-595 turned off
18 hours with N-595 operating
2 months, new fully charged battery
After 2 months storage the
N-595 will run for 50% of stated battery
life
91/157/EEC
5 ºC to 40 ºC (41 ºF to 104 ºF)
-390 m to 3,012 m
(-1,254 ft. to 9,882 ft.)
70 kPa to 106 kPa
(31.3 in. Hg to 20.6 in. Hg)