SATO CL4NX Operator's Manual page 97

Hide thumbs Also See for CL4NX:
Table of Contents

Advertisement

GAP Levels
Printing > Advanced > Calibrate > GAP Levels
Manually set the Gap sensor level.
The setting procedure is as follows:
First, adjust the "Low" level (voltage) of the Gap sensor.
1. Remove the label from the liner.
2. Pass the liner through the media sensor. Align it so that the media
sensor does not sense the I-mark (black mark).
3. Close the print head. To get the correct adjustment result, adjust after
you have closed the print head.
4. Select the GAP Levels in the Calibrate menu and press the
button.
/
5. Press the
buttons to change the Emit value until the Sensor
value is below 0.5 (V). Set the Emit value as low as possible.
6. If the Sensor value does not decrease below 0.5 after you changed
the Emit value, press the
7. Take a note of the Sensor value from the above procedure. This is the
"Low" level value for the Gap sensor.
Next, check the "High" level (voltage) of the Gap sensor as follows:
8. Pass the media (attached with liner) between the media sensors. Align
it so that the media sensor does not sense the I-mark (black mark).
9. Close the print head.
10. Check the Sensor value.
If the value is 1.0 (V) higher than the "Low" level value you have
recorded, then this is the "High" level value for the Gap sensor.
If the difference between the "High" and the "Low" levels is less than
1.0, adjust the Emit and Receive values so that the difference is more
than 1.0, or perform the adjustments again from step 1.
The standard values for the "High" and "Low" levels for the Gap sensor
are as follows:
• Low (with only liner) ≤ 0.5 (V)
• High (media attached with liner) - Low ≥ 1.0 (V)
11. If both "High" and "Low" levels comply with the standard value, press
the right soft button to confirm the value.
/
buttons to change the Receive value.
4 Operation and Configuration
CL4NX/CL6NX Operator Manual
95

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cl6nx

Table of Contents