Table E.8 Drift Specifications For 12000 Watt Units (80V To 600V Models) - Xantrex XDC 10-600 Operating Manual

6000 watt and 12000 watt series digital programmable dc power supply
Hide thumbs Also See for XDC 10-600:
Table of Contents

Advertisement

Table E.8Drift Specifications for 12000 Watt units (80V to 600V Models)
Models
1
Drift (30 minutes):
Voltage (0.04% of Vmax)
Current (0.6% of Imax)
2
Drift (8 hours):
Voltage (0.02% of Vmax)
Current (0.05% of Imax)
4
Temperature Coefficient:
Voltage (0.04% of Vmax/°C)
Current (0.06% of Imax/°C)
1. Maximum drift over 30 minutes with constant line, load, and temperature, after power on.
2. Maximum drift over 8 hours with constant line, load, and temperature, after 30 minute warm-up.
3. Current drift for 600V-20A unit is 0.08% of Imax.
4. Change in output per °C change in ambient temperature, with constant line and load.
Release 3.0
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
80–150
32 mV
900 mA
16 mV
3
75 mA
32 mV
90 mA
Specifications and Characteristics
Electrical Specifications—Summary
100–120
150–80
300–40
40 mV
60 mV
120 mV
720 mA
480 mA
240 mA
20 mV
30 mV
60 mV
60 mA
40 mA
20 mA
40 mV
60 mV
120 mV
72 mA
48 mA
24 mA
600–20
240 mV
120 mA
120 mV
16 mA
240 mV
12 mA
235