Calibration

Exit calibration mode

Exit calibration mode
When you have completed calibration, hit the EXIT key.
The SCPI command is:
CAL:STAT OFF, "0000"

Restore Factory Calibration

To restore the unit to the calibration constants set at the factory:
Step # Do This
1
2
3
The SCPI command is
CALibration:RESTore
Note This procedure should not be used in place of regular calibration, but may be useful to
restore the unit to an operational state in case of failure.
192
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
or
Scroll to select Factory Cal.
Select Y to Restore the factory
calibration. Selecting N will exit.
You Will See
Output V Cal
Factory Cal
Restore? Y
Operating Manual for XDC Series Power Supply