Remote Operation

Remote Analog Operation

Remote Analog Operation
Analog
The analog interface has 2 ports: the user lines and the analog programming lines.
Connections
The tables below show the function and power flow for each pin on these ports.
The user lines are optically isolated. The output lines are open collector
configuration. The input lines are capable of sinking 10mA (with recommended 5V
at input) up to a maximum 90mA.
The analog program and monitor lines are fully isolated from the supply output, but
not each other. Use precision variable low noise voltage sources for the program
lines, and be sure that the program source ground potential is the same as on the
analog readback circuitry.
The external analog monitoring circuitry must be high impedance because the
onboard V and I readback sources have approximately 300 ohms output impedance.
Connect your program and readback lines to the removable wire clamp connectors
marked "user lines" and "program lines" in Figure 4.1. Strip 0.2" (5mm) of
insulation from the wires and clamp securely at the appropriate pin. See Table 4.1,
Figure 4.1 and Table 4.2 for the location and function of each connector.
Use shielded twisted pairs of 22–24AWG for signal connections.
Pin Connections

Table 4.1 User Line Pins

Pin #
A1
A2
A3
A4
A5
A6
A7
108
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Function
Aux Status Line A
Aux Status Line B
External Trigger 4–12V
Safety Interlock
(Shutdown) 4–12V
Safety Interlock
(Shutdown) GND
User Power, 5–12Vdc
User Ground
Input/Output
Output
Output
Input
Input
Input
Input
Input