Xantrex XDC 10-600 Operating Manual page 193

6000 watt and 12000 watt series digital programmable dc power supply
Hide thumbs Also See for XDC 10-600:
Table of Contents

Advertisement

Calibration
Remote Interface Calibration Procedure
3. Enter voltage data
Enter the voltage at the current programming lines, read from the external DVM.
CAL:ANAL:5V:PROG:CURR:DATA <voltage>
4. Maximum calibration level
Set the input to the programming lines to approximately 4.5V (90% of full
scale).
Set the power supply to receive 5V analog current programming calibration data
by sending the command:
CAL:ANAL:5V:PROG:CURR:LEV MAX
5. Enter voltage data
Enter the voltage at the current programming lines, read from the external DVM.
CAL:ANAL:5V:PROG:CURR:DATA <voltage>
6. Power supply calculates and stores calibration constants.
Analog Programming Interface Current Readback Calibration
1. Attach a DVM across the current readback lines, Pins B6 and B1(GND).
2. Minimum calibration level
Set the current readback lines to approximately 10% of full scale by sending the
command:
CAL:ANAL:5V:READ:CURR:LEV MIN
3. Enter voltage data
Enter the voltage across the current readback lines, read from the external DVM.
CAL:ANAL:5V:READ:CURR:DATA <voltage>
4. Maximum calibration level
Set the current readback lines to approximately 90% of full scale by sending the
command:
CAL:ANAL:5V:READ:CURR:LEV MAX
5. Enter voltage data
Enter the voltage across the current readback lines, read from the external DVM.
CAL:ANAL:5V:READ:CURR:DATA <voltage>
6. Power supply calculates and stores calibration constants.