The slew rate error increases as the slew rate increases.
Selecting SLEW RATE from the main menu will give you two choices:
Voltage slew - adjust the voltage slew rate
Voltage default - restore the default voltage slew rate
Selecting VOLTAGE DEFAULT will return the slew rate to the default value of
1% rated voltage per 150us.
To set the slew rate:
Step # Do This
1
2
3
4
5
6
7
8
The combination must not exceed 1%/30us. Though the software will attempt to
achieve higher slew rates, it is limited to this value by hardware constraints. The
power supply slew rate cannot be faster than the rise and fall times given in the
product specifications on page 228. Manufacturer's recommendation is to keep the
slew rate at or below the default value.
Release 3.0
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
MENU
Use the scroll keys to get the SLEW
RATE menu.
ENTER
ENTER
Enter the desired voltage step.
Allowable range is 0.1% to 5% rated
voltage.
ENTER
Enter the time interval. The range is
150 us to 1.5 s.
ENTER
Power Supply Operation
You Will See
ERROR MSGS
SLEW RATE
Voltage slew
dV:
0.100 V
dV:
1.000 V
dt:
150 us
dt:
1000 us
Operation
103