Table E.2 Drift Specifications For 6000 Watt Units (10V To 60V Models) - Xantrex XDC 10-600 Operating Manual

6000 watt and 12000 watt series digital programmable dc power supply
Hide thumbs Also See for XDC 10-600:
Table of Contents

Advertisement

Table E.2Drift Specifications for 6000 Watt units (10V to 60V Models)
Models
1
Drift (30 minutes):
Voltage (0.04% of Vmax)
Current (0.6% of Imax)
2
Drift (8 hours):
Voltage (0.02% of Vmax)
Current (0.04% of Imax)
3
Temperature Coefficient:
Voltage (0.04% of Vmax/°C)
Current (0.06% of Imax/°C)
1. Maximum drift over 30 minutes with constant line, load, and temperature, after power on.
2. Maximum drift over 8 hours with constant line, load, and temperature, after 30 minute warm-up.
3. Change in output per °C change in ambient temperature, with constant line and load.
Release 3.0
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Specifications and Characteristics
10–600
20–300
30–200
4 mV
8 mV
12 mV
3600 mA
1800 mA
1200 mA
2 mV
4 mV
6 mV
240 mA
120 mA
80 mA
4 mV
8 mV
12 mV
360 mA
180 mA
120 mA
Electrical Specifications—Summary
40–150
60–100
16 mV
24 mV
900 mA
600 mA
8 mV
12 mV
60 mA
40 mA
16 mV
24 mV
90 mA
60 mA
229