Teledyne API Model 401 O

1.4.10 Flow Rates

800 scc/min ± 10% (standard cubic centimeters/minute) sample flow rate.
2-5 l/min ± 10% (standard cubic liter/minute) O

1.4.11 Power

Standard input power: 120VAC rms, 60 Hz, 250 watts maximum and 220, 240 V, 50 Hz.
5A resettable circuit breaker.

1.4.12 Weight

27 Kg (60 lb.) maximum.

1.5 Installation And Overview

The Model 401 is shipped with the following standard equipment:
1. Power cord.
2. Instruction manual.
To avoid personal injury, always use two persons to lift and carry
Upon receiving the Model 401 please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please advise shipper first,
then Teledyne API.)
2. Remove all red colored shipping screws from the underside of the instrument. Note: Save
these shipping screws and re-install them whenever the unit is shipped to another location.
3. When installing the Model 401, allow a minimum of 4 inches (100mm) of clearance at the
back of the instrument and 1 inch (25mm) of clearance on each side for proper ventilation.
4. Check that all options ordered are enclosed.
5. Connect analyzer' sample inlet line to the M401 sample port of manifold on rear panel.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
3
CAUTION
the Model 401.
1-7