Errors appearing on pop-up menus

error message
description
affected modules
action
error message
description
affected modules
action
error message
description
affected modules
action
error message
description
affected modules
action
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com
Errors appearing on pop-up
menus

Error on Module

Calibration data corrupted
The calibration data loaded on the module is invalid. Either the caldata version
number is not compatible with the firmware or the caldata is corrupt (checksum
incorrect)
all modules
Please return the unit to your local service center for repair
Laser safety error
The laser control circuit has switched of the laser. Either the laser current or the
laser power was too high.
TLS
Try to run the automatic realign from the menu or to operate the laser at a lower
power level. If the failure persists, please return the unit to your local service
center for repair.
Temperature out of range
The temperature to operate the tunable laser source module was measured
outside given limits.
TLS
Verify the fan is working and the ventilation system is not blocked. If the failure
persists, please return the unit to your local service center for repair.
Zeroing failed. Light?
Zeroing has still detected some light. Either the input is not darkened properly
or there is some error on the powermeter module/head.
Powermeter
Darken the optical input(s) properly. If the failure persists, please return the unit
to your local service center for repair.