Figure 2-36 Recorder Maximum Pop-Up; Table 2-4 Ranges Of Recorder Output Setting - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

2
General Power Meter Functions

Figure 2-36 Recorder Maximum pop-up

11 Use the numeric keys to enter the power level you want to generate a 0
12 Press
The recorder output can be disabled and re-enabled by checking the Rec o/p: setting field.
N O T E
The highest power you are going to measure is used to determine the value which you
should set for the Recorder Output maximum setting. For example, if you are measuring a
power less than 1 mW and greater than 100 mW, then set the recorder maximum value to
1 mW.

Table 2-4 Ranges of recorder output setting

Log
Lin
Log
Lin
74
Vdc output in the Recorder Minimum pop- up and press
key to close the Measurement Setup screen.
50
40
100 W
10 W
–10
–20
100 mW
10 mW
30
20
1 W
100 mW
–30
–40
1 mW
100 nW
N1913/1914A EPM Series Power Meters User's Guide
dBm
.
10
0
10 mW
1 mW
–50
–60
10 nW
1 nW