7.2 Safety summary

7.2.1 Insulation resistance (IR)

Referring to the working condition, the insulation resistance of the power terminal and the case
should not be less than 50MΩ.
Under the damp and heat condition, the insulation resistance should not be less than 2 MΩ.

7.2.2 Insulation intension

Referring the working condition, the power terminal and the case should load the AC voltage with
50Hz in frequency and 1.5kV in rated voltage for 1 minute, and should have no breakdown and
arcing.

7.2.3 Leakage current

The leakage current should not be more than 3.5mA (The AC effective value).

7.2.4 The EMC requirement

The ransient sensitivity of the instrument power meets the GB6833.4.
The conduction sensitivity of the instrument meets the GB6833.6.
The radiated interference of the instrument meets the GB6833.10.
STB6684
4 bins: (3 bins for PASS, 1 bin for FAIL)
Auto, Hold
1 to 100
20 sets of setup values can be stored.
GPIB (optional), RS232C,
HANDLERinterface,
USBDEVICE( USBTMC and USBCDC support),
USBHOST(FAT16 and FAT32 ), SCANNING