STB6684/A Operation Manual
Operational steps for switching on or off open correction:
(1)Move the cursor to OPEN CORR, the following softkeys will be displayed:
ON
This softkey is used to switch on the open correction function.
OFF
This softkey is used to switch off the open correction function.
(2) Use above softkeys to switch on or off the open correction.

4.8.2 OPEN CORR

This function corrects single range or all ranges.
The following ranges can be corrected:
1mA
: Correct in the range of 1mA.
100uA
: Correct in the range of 100μA.
10uA
: Correct in the range of 10μA.
1uA
: Correct in the range of 1μA.
100nA
: Correct in the range of 100nA.
10nA
: Correct in the range of 10nA.
1nA
: Correct in the range of 1nA.
ALL
:
Correct all ranges one times.
Operational steps for setting open correction:
(1)Move the cursor to a range below OPEN CORR, the following softkeys will be
displayed:
APPLY
This softkey is used to execute the open correction in a specific range.
When the correction is successfully done, [YES] will be displayed behind the
range; or [NO] will be displayed.
CLEAR
This softkey is used to clear the open correction data in a specific range.
When the clear operation is successfully done, [NO] will be displayed behind
the range; if the correction data is also present, [YES] will be displayed.
(2) Use above softkeys to set the open correction.
Chapter 4 [SETUP]
◇ 59