2) Press a softkey to select the desired sequence. If the desired sequence is not displayed in
the current display page, press MORE 1/2 to display the following softkeys.
a)
USER1
b)
USER2
c)
USER3
d)
USER4
e)
MORE 2/2
3) If the desired sequence is missed, repeat step1 and step 2 until the desired sequence is
selected.

4.4.2 DEFAULT SEQ.

Parameters of default sequence can be scanned or copied but cannot be changed.
Including: 20 SEC SEQ.
60 SEC SEQ.
20 SEC MEAS. GO
60 SEC MEAS. GO
Introduction to default sequences:
(1)When one of the first two default sequences is selected, the meaure will executed in
continuous mode defined in the <SEQUENCE SETUP> page. At 20s or 60s, the measure will
stop and DUT will automatically discharge.
(2)When "Measure to Go" is selected, the measure will stop and DUT starts discharging as
soon as the low limit is met. At 20s or 60s, if the low limit cannot be met, the measure will
stops like that happens at 20 SEC SEQ. or 60 SEC SEQ..
Operational steps for setting the default sequence:
1) Move the cursor to DEFAULT SEQ., the following softkeys will be displayed:
EDIT
Press EDIT to enter into the <SEQ. CONTENT> page.In this page, only high and low
limits can be set, see 4.5.7 HIGH, LOW. Other parameters can only be scanned.
COPY
Press this key, the instrument will copy all parameters in the current page to the memory,
but only the latest contents can be copied.
Chapter 4 [SETUP]
◇ 39