Bar Scale - Saluki STB6684 Operation Manual

Insulation resistance meter
Table of Contents

Advertisement

STB6684/A Operation Manual
————————————————————————————————————
NOTE: In <BARGRAPH DISP>, pressing [TEST], the instruement will start or stop a test;
————————————————————————————————————

3.3.1 BAR SCALE

STB6684/A has two bar scale settings: resistance and current, as follows:
Press [SETUP] and select RESULT, CURR should be chosen in the softkey zone. Move the
cursor back to <MEASURE SETUP> and enter into the <BARGRAPH SETUP> page,
the bar scale of the current range can be set.
When STB6684/A is in CURR mode, the setting range ranges from 0 to 8.
0 : AUTO SCALE
1 : 1pA – 100pA
2 : 10pA – 1nA
3 : 100pA – 10nA
4 : 1nA – 100nA
5 : 10nA – 1uA
6 : 100nA – 10uA
7 : 1uA – 100uA
8 : 10uA – 1mA
Press [SETUP] and select RESULT, RES should be chosen in the softkey zone. Move the
cursor back to <MEASURE SETUP> and enter into the <BARGRAPH SETUP> page,
the bar scale of the resistance range can be set.
When STB6684/A is in RES mode, the setting range ranges from 0 to 10.
0 : AUTO SCALE
1 : 10kΩ - 1 MΩ
2 : 100kΩ - 10 MΩ
3 : 1MΩ - 100 MΩ
4 : 10MΩ - 1GΩ
5 : 100MΩ - 10GΩ
6 : 1GΩ – 100GΩ
7 : 10GΩ – 1TΩ
8 : 100GΩ – 10TΩ
9 : 1TΩ – 100TΩ
10 : 10TΩ – 1PΩ
pressing [HV], it will start or stop the high voltage output.
(automatically select scale)
(automatically select scale)
◇ 21
Chapter 3 [DISP]

Advertisement

Table of Contents
loading

This manual is also suitable for:

Stb6684a

Table of Contents