I1_Max; Hv2 - Saluki STB6684 Operation Manual

Insulation resistance meter
Table of Contents

Advertisement

STB6684/A Operation Manual
This is a coarse adjustment key to decrease the voltage in accordance with the
sequence of some common-used voltage points.Take STB6684A for example,
the
common-used
10V,50V,100V,250V,500V, 1000V,1005V.
2) User can use the softkeys to adjust the value of HV1 or the numerical keys to input the
desired value of HV1. Voltage units (V) currently available will be displayed in the
softkey zone when the numerical keys are used to input the voltage. Pressing the
corresponding softkey, the voltage unit can be selected and the data will be confirmed.
When [ENTER] is pressed to finish the data input, the default voltage unit is V.

3.1.2 I1_MAX

This function is used to set the maximum charge current allowable by HV1. Three currents
are selectable: 2mA, 25mA and 200mA.
Operational steps for the I1_MAX setting:
1)
Move the cursor to I1_MAX, the following items will be displayed in the softkey
zone:
2 mA
25 mA
200 mA
2)
Press one of above softkeys to set the charge current.

3.1.3 HV2

HV2 output is available only on STB6684 but not on STB6684A.
This function is used to set the voltage (preset voltage) output of HV2. There are two
voltage settings available: voltage independent mode (HV2 TRACK: OFF) and voltage
track mode (HV2 TRACK: ON).
<1> Voltage setting in voltage independent mode( HV2 TRACK: OFF):
Voltage setting ranges:
STB6684 : 10V – 505V
HV2 is not available on STB6684A.
voltage
points
◇ 13
Chapter 3 [DISP]
are
as
follows: