STB6684/A Operation Manual
AUTO
: auto range. In range of 1nA to 10nA, the input internal impedance is 1MΩ;
and in other ranges, it is 10kΩ.
1mA
: 10OuA – 1mA ;
100uA
: 10uA – 100uA ; input internal impedance 10 kΩ
10uA
: 1uA – 10uA ;
1uA
: 100nA – 1uA ;
100nA
: 10nA – 100nA ; input internal impedance 10 kΩ
10nA
: 1nA – 10nA ;
1nA
: 10pA – 1nA ;
Operational steps for setting the range in the sequence content page:
1)Move the cursor to RANG, the following softkeys will be displayed:
↑(+)
This softkey is used to increase the range in accordance with the sequence listed
above.
↓(-)
This softkey is used to decrease the range in accordance with the sequence listed
above.
2)Select the corresponding softkey to set the range.

4.5.6 AVG

Setting range: 1 – 100.
Operational steps for setting the average times in the sequence content page:
1)Move the cursor AVG, the following softkeys will be displayed:
↑(+)
Use this key to increase the average times.
↓(-)
Use this key to decrease the average times.
2) Use above softkeys to select the average times or numerical keys and [ENTER] to input the
desired average times.
input internal impedance 10kΩ
input internal impedance 10 kΩ
input internal impedance 10 kΩ
input internal impedance 1MΩ
input internal impedance 1MΩ
◇ 49