Limit Setup - Saluki STB6684 Operation Manual

Insulation resistance meter
Table of Contents

Advertisement

STB6684/A Operation Manual

4.2 LIMIT SETUP

Press [SETUP], and LIMIT SETUP to enter into the <LIMIT SETUP> page.
Figure 4-2:
Comparator function and 4 bin limits can be set on this page. The result can be divided
into up to 5 bins (from BIN1 to BIN5). If STB6684 has installed a HANDLER interface and
is used in an automatic test sorting system, the comparator function will be quite
useful. Limit parameters in comparator function can be set only in the <LIMIT SETUP>
page.
————————————————————————————————————
NOTE: Comparator function is only usable in <MEASURE DISP>, <BARGRAPH DISP>
and <BIN DISP> pages.
Programmable parameters:
Comparator function: ON/OFF (COMP)
Measure parameters(PARAM)
Nominal value(NOM)
Comparator function limit mode(MODE)
Low limit of each bin(LOW)
High limit of each bin(HIGH)
Chapter 4 [SETUP]
◇ 31