STB6684/A Operation Manual
Command syntax:
TRIGger:MODE
For example: WrtCmd("TRIG:MODE CONT"); set the trigger mode to CONTinue.
Query syntax: TRIGger: MODE?
Return format:
CONTINUE
SINGLE
9.5.5 HTOUtput subsystem commands
HTOUtput subsystem commands set the high voltage, in CONTINUE mode, to ON or OFF.
Command syntax:
HTOUtput
For example: WrtCmd("HTOUtput ON");
Query syntax: HTOUtput?
Return format:< NR1> <NL^END>
9.5.6 HUMReject subsystem commands
HUMReject subsystem commands set the instrument working frequency.
Command syntax:
HUMReject
For example: WrtCmd("HUMReject 50 ");
Query syntax: HUMReject?
Return format:
9.5.7 CCHEck subsystem commands
CCHEck subsystem commands set the contact check function to ON or OFF.
Command syntax:
CCHEck
For example: WrtCmd("CCHEckON");
CONTinue
SINGle
+ <NL^END>
ON (1)
OFF (0)
50 (50Hz)
60
(60Hz)
50Hz
<NL^END>
60Hz
ON (1)
OFF (0)
◇ 112
Chapter 9 Command Reference