STB6684/A Operation Manual
4.8.3 LOAD CORR: ON/ OFF
LOAD CORR: ON
LOAD CORR: OFF Do not take the open correction value into account.
Operational steps for switching on or off the loac correction function:
(1)Move the cursor to LOAD CORR, the following softkeys will be displayed:
■
ON
This key is used to switch on the load correction.
■
OFF
This key is used to switch off the load correction.
(2) Use above softkeys to switch on or off the load correction.
4.8.4 REFERENCE
In this zone, the correction reference value can be input, and the load correction can be
enabled and cleared.
Operational steps for inputting the reference value:
Move the cursor to REFERENCE, and input the desired data. In the softkey zone, units pA,
nA, µA, mA or kΩ, MΩ, GΩ, TΩ, PΩ will be displayed, press a key to select the desired
unit and end inputting. When [ENTER] is used to finish inputting, the default unit is A or Ω.
Operational steps for setting reference:
(1)Move the cursor to REFERENCE, the following softkeys will be displayed:
START
■
In load correction mode, this key is used to start up HV and test, and to display
voltage, current and resistance.
APLLY
■
This key is used to confirm, record and calculate the load correction coefficient.
CLEAR
■
This softkey is used to clear the load correction coefficient.
■
STOP
This softkey is used to turn off the voltage and to end the test.
(2) Use above softkeys to set the reference.
4.8.5 FILE
See 5.2 for details about the <FILE MANAGE> function.
Take the open correction value into account.
Chapter 4 [SETUP]
◇ 60