Limit Subsystem Commands - Saluki STB6684 Operation Manual

Insulation resistance meter
Table of Contents

Advertisement

STB6684/A Operation Manual

9.5.10 LIMIt subsystem commands

LIMIt subsystem commands set bin comparator functions such as comparator ON/OFF, limit list.
Command tree:
LIMIt
The [:STATe] command sets the comparator function to ON or OFF.
The :LIMIt [:STATe]? query returns the current status of the comparator funtion.
Command syntax:
LIMIt [:STATe]
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd("LIMIt ON")
Query syntax: LIMIt [:STATe]?
Return format: <NR1><NL^END>
The :MODE command sets the limit mode for the comparator function.
The :LIMIt:MODE? query returns the current limit mode.
Command syntax:
LIMIt:MODE
Where,
ATOLerance
PTOLerance
SEQuence
For example: WrtCmd("LIMIt:MODE ATOL")
Query syntax: LIMIt:MODE?
Return format:
ATOL
PTOL
[:STATe] ON (1)
OFF(0)
:MODE Absolute TOLerance
Percent TOLerance
SEQuence
:TOLerance
:NOMinal <value>
:BIN<n> <low limit>,<high limit>
:SEQuence
:BIN <BIN1 low limit>,<BIN1 high limit>,
:PARAM
Res
Cur
ON
OFF
1
0
ATOLerance
PTOLerance
SEQuence
Set the limt mode to ATOLerance.
Set the limit mode to PTOLerance.
Set the limit mode to SEQuence.
<NL^END>
<BIN2 high limit>,<BIN3 high limit>,
...... ,<BINn high limit>
◇ 114
Chapter 9 Command Reference

Advertisement

Table of Contents
loading

This manual is also suitable for:

Stb6684a

Table of Contents