STB6684/A Operation Manual

9.5.4 TRIGger subsystem command

TRIGger subsystem commands are used to set the intrument trigger source and the delay time after
triggering and to trigger measurements.
Command tree:
TRIGger
The IMMediate command is used to trigger or switch off test.
Command syntax:
TRIGger[:IMMediate]
For example: WrtCmd("TRIG ON");
————————————————————————————————————————————
NOTE: TRIGGER ON/OFF commands in <MEAS DISP> and <BIN DISP> are the same.
Input the command first time, the instrument take a test; Input the command second
time, the instrument will stop testing.
————————————————————————————————————————————
The :SOURce command sets the trigger source mode.
The :TRIGger:SOURce? query returns the current trigger source mode.
Command syntax:
TRIGger:SOURce
Where,
EXTernal
BUS
HOLD
For example: WrtCmd("TRIG:SOUR BUS");
Query syntax: TRIGger:SOURce?
Return format:
The :MODE command sets the intrument trigger mode.
The :TRIGger: MODE? query returns the current trigger mode.
[:IMMediate] ON
OFF
:SOURce EXTernal
BUS
HOLD
:MODE CONTinue
SINGle
ON
OFF
EXTernal
BUS
HOLD
Triggered by HANDLER interface.
Triggered by RS232 interface or GPIB interface.
Triggered by pressing Test.
EXT
<NL^END>
BUS
HOLD
◇ 111