STB6684/A Operation Manual
Operational steps for the trigger mode setting:
The following operations can set the trigger mode (except BUS trigger mode). When the
BUS trigger mode need to be set, a "TRIGger:SOURce BUS"command will be sent to the
instrument via IEEE 488 interface.
1)
Move the cursor to TRIG, the following items will be displayed in the softkey zone.
MAN
EXT
BUS
2)
Above softkeys can set the instrument trigger mode.
————————————————————————————————————
NOTE: In EXT mode, the minimum current range is 10mA and cannot be switched to 1nA.
In MAN or BUS mode, the minimum current range can be switched to 1nA.

3.1.10 MODE

STB6684/A has 2 test modes: SINGLE and CONTINUE.
In SINGLE mode and by pressing [TEST] one time, the instrument makes a test in the
following sequence of trigger, charge, wait, test and end test.
In CONTINUE mode and by pressing [TEST], the instrument will make tests continuously;
press [TEST] again, the instrument will exit the continue test mode.
Operational steps for the test mode setting:
1)Move the cursor to MODE, the following items will be displayed in the softkey zone:
SINGLE
CONTINUE
2)Above softkeys are used to set the instrument test mode.

3.1.11 FILE

See 5.2 for details about the <FILE MANAGE> function.