STB6684/A Operation Manual
This softkey is used to decrease the scan number by 1.
2)Use above softkeys to set the scan end number, or numerical keys and [ENTER] to input
the desired one.
————————————————————————————————————
NOTE: The scan start number must be less than the scan end number.
————————————————————————————————————

4.7.4 SCAN RESULT

This zone is used to set the result display as MEASUREMENT or PASS/ FAIL.
Operational steps for setting the scan result display mode:
(1) Move the cursor to SCAN RESULT, the following softkeys will be displayed:
MEASUREMENT
This softkey is used to set the display mode as measurement.
PASS/ FAIL
This softkey is used to set the display mode as PASS/ FAIL.
(2)Use above softkeys to set the scan result display mode.

4.7.5 SCAN CONNECT

This zone is used to check the scanner is connected correctly or not.
Operational steps for setting scan connect:
(1)Move the cursor to SCAN CONNECT, the following softkey will be displayed:
MEASURE
This key is used to start up the measurement that whether the scanner is connected
correctly or not.
(2) Use the above softkey to start up measure.
If YES is displayed, the scanner is connected correctly.
If NO is displayed, the scanner is not connected correctly.

4.7.6 FILE

See 5.2 for details about the <FILE MANAGE> function.
Chapter 4 [SETUP]
◇ 57