4.5.7 LOW, HIGH

Low and high limits in the sequence content page can be set to current or resistance limits,
which depends on [RESULT] in the <MEASURE SETUP> page. An exceptional case is in
FLASH TEST, it can only be set to current limits.
If [RESULT] displays CURRENT, it can only be set to current limits, on the contrary is
resistance limits.
————————————————————————————————————
NOTE:
1. User can decide to set the HIGH/ LOW limit or not.
If both limits or either one are set, STB6684/A will automatically switch on the compare
function.
Likewise, if both limits are not set, the compare function will be automatically switched
off.
2. In test modes of MEASURE and MEAS. CONT, HIGH/ LOW limits are not necessary.
3. In test modes of MEAS. TO GO, FLASH TEST, HIGH/ LOW limits must be set or the test
will terminate and an error prompt will be displayed.
————————————————————————————————————
Operations steps for setting LOW, HIGH limits int the sequence content page:
1) Move the cursor to LOW and UPP, the following softkeys will be displayed:
If the limit mode is in current mode:
■ pA
■ nA
■ uA
■ mA
If the limit mode is in resistance mode:
■ kΩ
■ MΩ
■ GΩ
■ TΩ
■ PΩ
2)Use numerical keys and above softkeys to input low and high limits.
————————————————————————————————————
NOTE: No matter in current mode or resistance mode, the low limit must be less than the
Chapter 4 [SETUP]