STB6684/A Operation Manual
mA) or (kΩ, MΩ, GΩ, TΩ, PΩ) to replace [ENTER] to input the limit unit. When use
[ENTER] to end inputting, the default unit is the same as that input last time.
7)
After inputting the low limit value of bin1, the cursor is jumped to HIGH of bin1. Input
the high limit of bin1.
8)
The cursor is jumped to HIGH of bin2. As the limit mode of compare function is
sequential mode, the low limit of bin2 is equal to the high limit of bin1. Input the high
limit of bin2.
9)
The cursor is jumped to HIGH of bin3. Input the high limit of bin3.
10) The cursor is jumped to HIGH of bin 4. Input the high limit of bin4. The settings for
high and low limits are done.
4.2.5 File manage
See 5.2 for details about the <FILE MANAGE> function.
4.2.6 TOOLS
The clear table function can clear all limit parameters.When the limit mode is changed, user
must use the clear table function to clear all limit parameters shown in the current page.
Operational steps for clearing limit parameters in the <LIMIT SETUP> page.
1)
Move the cursor to TOOLS, the following softkey will be displayed:
CLEAR TABLE
2)
Press this softkey to clear all limit parameters in the current page.
4.3 BARGRAPH SETUP
Press [SETUP], and then BAR. SETUP to enter into the <BARGRAPH SETUP> page.
Figure 4-3:
Chapter 4 [SETUP]
◇ 35