10.3 BIN signals

10.3.1 BIN signals in the test display page.

Limits
(ascending)
LIM0
Bin Out
BIN0
Resistance mode
Limits
(ascending)
100Gohm
Bin Out
BIN0
"Bad"
Current mode
Limits
(ascending)
5nA
Bin Out
BIN0
"Good"
Above figure shows the layout of BIN signals that is in accordance with limit setup. If the test
value is larger than LIM1 and less than LIM2, the BIN2 signal will be activated (the output
optocoupler will be switched on).
Both in resistance mode (figure 9.3) and current mode (figure 9.3), limit setup is organized in
ascending order.This means that when in resistance mode and a LIM value is set, a "goood"
component will set as BIN5; while in current mode, it will be set as BIN0. More limit setups can be
used to exclude shorted or open components.
_______________________________________________________________________________
NOTE:
The instrument must be set to resistance mode when resistance limit setup is to be used.
The instrument must be set to current mode when current limit setup is be used.
_______________________________________________________________________________
LIM1
LIM2
BIN1
BIN2
BIN3
BIN1
BIN2
BIN3
BIN1
BIN2
BIN3
Figure 9.3 Layout of BIN signals
◇ 132
LIM3
LIM4
BIN4
BIN5
BIN4
BIN5
"Good"
BIN4
BIN5
"Bad"