4.2.6. Controls used for Input/Output
Fig.4.9.
DIGITAL 1/0:
r - - - D i g i t a l
1/
0 - - - - - - .
Input
Output
Plot
Stop
Local
IIIII
a a a
1
.
o..:olc,.
.• ,.., ...
t$'1)1"•l""•l••••
fypt
2f}~}
·
r ;;;;;
~
a
11
a
a
!l
(it
:'i.'t""·~-
....
..
,.
e., . .
Controls used for Input/Output on the 2032
"Input" -
causes the 2032 to input "Total Documentation" and other
files over its digital interface under manual control. The "Input" L.E.D.
comes on, indicating that the 2032 is a listener.
The 2032 Display is updated, where necessary, after each file is input,
whereby the next file can be read. On completion of the digital input,
DIGITAL 1/0 "Stop" may be manually selected, whereby the "Input"
L.E.D. goes off, indicating that the 2032 is no longer a listener.
Total Documentation files holding averaged data can, depending on
the setting of the MEMORY "Enable" pushkey, (see Section 4.2.2), be
read into the input buffer or store buffer. When MEMORY "Enable" is
on, (that is, L.E.D. on), read-in will take place to the store buffer, and
when it is off, (L.E.D. off) to the input buffer.
For details of the formats required by Total Documentation and other
files for digital input to the 2032, refer to the separate 2032 IEC Inter-
face Manual.
"Output" -
allows digital output of Total Documentation files con-
sisting of both basic measured data, (input or stored), and the MEA-
SUREMENT SETUP, DISPLAY SETUP, CURSOR SETUP, and the text-
string, under manual control. The measured data output is:
Single Spectrum Averaging Ch. A or 8
2048 sample Time Function
1024 line Autospectrum
Single Spectrum Averaging Zero Pad Ch. A or 8
1024 sample Time Function
1024 line Autospectrum
41