BRUEL & KJAER 2032 Instruction Manual page 140

Dual channel signal analyzer
Table of Contents

Advertisement

12.3.2. Sine Generator Frequency
2. SINE GENERATOR FREQ.
*
Resolution: 2--a Hz
=
15,625 mHz
0,000 Hz*
0,015 Hz
0,031 Hz
25559,984 Hz
25600,000 Hz
If external sampling is selected, then the Sine Generator Frequency line becomes:
I
I* c
SINE GENERATOR FREQ.
(Frequency)
where C
= [
(external sampling frequency) I 65536 ] Hz
12.3.3. Special Parameter
130
3. SPECIAL PARAMETER
#
1.___ _ _
-..J
0
1
112
113
(Parameter name)
(
Dec.}
Oct.}
VALUE:
{
- ~2768
32767
{
10~000
077777
On power on, SYSTEM RESET 2, and SYSTEM RESET 9, all Special Parameter values
are set to 0 meaning that default values are used, ensuring normal operation of the
2032. Special Parameter 0, Main Key, acts as a key to the other Special Parameter
values. Access to the other Special Parameter values can only be obtained by keying
in one of the following four numbers into the value field of Special Parameter 0:
203210
2032
8
203410
2034
8
(=
3760
8 )
( =
1 050 10 )
(=
3762
8 )
( =
1
052 10 )
(The numbers in brackets indicate what will be shown by ·the value field should a
decimal number be keyed in when the field is set to display in octal, and vice versa).
Note that while use of Special Parameter 0 is necessary to gain access to the other
Special Parameter values, these values can always be displayed. For further details of
the other Special Parameters, refer to Sections 12.4 to 12. 7.