5.5.
OTHER FIELDS ON THE 2032 DISPLAY
.
5.5.1. Data Source
Data Source
Fig.5.29.
The Data Source Control Field
The Data Source control field controls whether the displayed data is taken from the
2032 input buffer or from the store buffer, or whether the equalized function is shown.
It is a hidden field which only becomes visible when it is set to Stored or Equalized or
when the Field Selector is positioned in the top line of the Display.
0.
INPUT
1. STORED
2. EQUALIZED*
*
Can only be selected when the Function field is set to Frequency Response or Impulse Response. A
measurement must have been stored before the Data Source field is switched to Equalized
.
If
Equalized is selected when the input measurement is measured with a different MEASUREMENT
SETUP to the stored
measurement,
a star will appear in the field as a warning. The equalized
function
is
a complex division of the input measurement by the stored
measurement.
89