Storage And Recall Of Display Setups - BRUEL & KJAER 2032 Instruction Manual

Dual channel signal analyzer
Table of Contents

Advertisement

3.5.1 Storage and Recall of DISPLAY SETUPS
The 2032 stores 22 DISPLAY SETUPS named Wn, Wn, 1, 2, 3, .... , 19, 20, (n
=
1, 2, 3, ... ,
19, 20). The DISPLAY SETUP(s) in use are identified in the DISPLAY SETUP label
field(s) at the head of the DISPLAY SETUP(s). All of the 22 setups are stored when
electrical power to the 2032 is switched off. Setups 1 to 10 are user-defined, and are
stored using the SAVE SETUP push keys, see Section 4.2.3. Setups 11 to 20 are factory
defined and held in read-only memories, see the Chapter 10. Setups 1 to 20 can be
recalled by pressing SETUP LABEL "Display", ensuring that with a dual-trace format
the Field Selector is placed in the label field of the DISPLAY SETUP to be changed,
entering the number of the required setup via the FIELD ENTRY numeric keyboard, and
pressing "ENT", or by pressing SETUP LABEL "Display" and using the FIELD ENTRY
knob or increment/decrement pushkeys. As an aid to selection, the setups can be
listed, (Fig.3.6), by pressing SETUP LABEL "Display" and then pressing the upper right
hand (blank) key of the FIELD ENTRY numeric keyboard. This key can then be used to
toggle the Display between a listing of the DISPLAY SETUPs and the selected MEA-
SUREMENT SETUP. Should a recalled setup be changed, it becomes a Working Setup,
labelled Wn, where n indicates the DISPLAY SETUP from which it is derived. There are
two Working Setups, one for each of displayable traces, but no Documentation or
Stored Setup. The numeric keyboard entry codes for the two Working Setups are 0
and -1.
Fig.3.6.
A listing of the available DISPLAY SETUPs on the 2032
Display
21