Dual Signal Enhancement:
2 x 2048 sample Time Function
2
x
2048 sample Enhanced Time Function
Dual Amplitude Probability:
2
x
2048
sample Time Function
2
x
512 class Amplitude Probability Density
All functions can be calculated and displayed
from the stored basic measurement data. The
measurement setup used Is stored together
with the data
POSTPROCESSING:
Equalize: Calculates the complex ratio of the
measured and stored Transfer Function. Ap-
plies also to Impulse Response
A-Weighting: Autospectrum, Coherent and
Non-coherent Output Power and Sound In-
tensity can be A-weighted in any frequency
span
Integration and Differentiation: Frequency
domain functions: Single and double division
or multiplication with jw can be selected.
Time domain functions: Single or double Inte-
grated or differentiated functions obtained by
inverse transformation from the frequency
domain
Delay Compensation: For compensating
Frequency Response phase measurements
for system delays. Adjustable from - T to
T.
Resolution
~
T 1128
Bow-Tie Correction: Correlation Functions
can be displayed with or without bow-tie
correction
Cepstrum Littering: Shortpass or longpass
with selectable length
SYSTEM ACCURACY:
Frequency Response: ± 0,4 dB at filter cen-
tres with DC coupling
Amplitude Linearity: ± 0,1 dB or ± 0,01% of
maximum Input voltage, whichever Is greater,
measured In Autospectrum with at least 32
Independent averages, In the presence of an-
other In-band signal not less than 20 dB below
max. Input
Dynamic Range:
>
75 dB, for Autospectrum
.
with at least 32 independent averages. Noise,
distortion, and spurious signals at least 75 dB
below max. Input voltage
.Minimum Detectable Signal Level: 1JLV
Frequency Accuracy and Stability: 0,01%
without warm-up, (no adjustment necessary)
OPERATION:
Measurement Setup: 10 complete user-de-
fined measurement setups can be saved and
recalled from a non-volatile memory. Mea-
surement setup parameters can be changed
before or during a measurement. Additionally
10 factory-defined complete measurement
setups are saved In ROM
Display Setup: 10 complete user-defined
display setups can be saved and recalled
from a non-volatile memory. Display setup
parameters can be changed before, during or
after a measurement. Additionally 1 0 factory-
defined complete display setups are saved In
ROM
Parameter Entry: Via Increment/decrement
controls, continuously variable speed field
entry control, or numeric keypad
Auto Scale: Display autoscale to maximize
the area of the CRT used. The largest numer-
lc value In the function on the display Is auto-
matically selected as new Y -full scale
Averaging Control: Start, Proceed, Stop
Start: Clears the average accumulator and
starts an average
Stop: Stops the averaging process or rejects
an estimate in manual accept
Proceed: Continues an average without
clearing the average accumulator or accepts
an estimate In manual accept
·
SIGNAL GENERATOR:
Impulse: Length:
~
T. Repetition: T. Ampli-
tude:
+
SV
Reference Sine: 5 V RMS. Frequency equals
selected centre frequency In zoom mode, and
50% of selected frequency span in baseband
mode
Random Noise: Random noise signal bandll-
mlted to match the selected frequency span.
Power spectrum flat within
±
2 dB In the se-
lected
span.
Output level: 90dBJLV/'{'i;F, RMS
Pseudo Random Noise: Pseudo random
noise signal with spectral line spacing match-
ing the line spacing
~
F for the selected fre-
quency span. The noise spectrum Is bandli-
mlted to match the selected frequency span.
Power spectrum flat within
±
2 dB In the se-
lected span. Output level: 90dBJLVI'{'i;F, RMS
Variable Sine: 5 V RMS. Frequency range:
15mHz
to
25,6kHz.
Setting
resolution:
15mHz
Output Impedance: 50
n
Level: Cal. or 0 to 20 dB attenuation
PLOT:
Graphics Recorder Type 2313: Any display
on 2032 Including graphics and all annotation
can be plotted on 2313 In less than 10 sec-
onds with any Application Package In 2313
X-Y Recorder Type 2308: Plot of single or
dual graph with axes. X: 0 to
+
10V. Y: 0 to
+
10 V. X and Y outputs are controlled for
constant writing speed with a choice of S
different writing speeds. Pen lift control.
Analysts continues during read-out
Video Output: For video hard copy units and
recorders. Output Impedance: 7S
n
MASS STORAGE:
Transmission to Digital Cassette Recorder
Type 7400 of flies containing basic measured
data (current or stored), Including measure-
ment setup, display setup, cursor setup and
text string. Later re-entry to 2032 allows re-
creation of any function. Also storage of
user-defined measurement and display set-
ups and of abbreviated read-outs
IEC/IEEE INTERFACE:
Conforms to IEC 62S-1 and IEEE 488 stan-
dards. Connection into an IEC interface sys-
tem Is made using cable AO 0194. Connec-
tion into an IEEE 488 Interface system is
made using cable AO 0194 and adaptor AO
019S and cable AO 0194
IEC/IEEE Function Implemented:
Source Handshake
SH1
Acceptor Handshake AH1
Talker
TS
Listener
L3
Service Request
SR1
Remote/Local
RL 1
Device Clear
DC 1
Data: Any function shown on display, any ba-
ste measured data, measurement setup, dis-
play setup, cursor setup, text string etc. can
be transmitted to and from 2032
Remote Control: All functions and Instru-
ment settings excl. Signal Generator Ampli-
tude and Microphone Polarization Voltage
can be remotely controlled
Command Set: Simple and easy to remem-
ber standard engineering English. Resistant
to operator error. Syntax errors displayed on
screen
Graphics: Display useable for multlmode
graphics
Code: ISO 7 -bit code or Binary
Interface Terminator: Can be defined from
the front panel or from a controller
Device Address: Displayable on screen
POWER SUPPLY:
100- 127V or 200- 240V AC ± 10%, 50-
60Hz± S%
Consumption approx. 400 VA
GENERAL:
Safety: Complies with IEC 348 Safety Class II
EMI: Complies with U.S. FCC requirement for
Class B computing devices
Operating Temperature:
soc
to 40°C (41 °F
to 104°F)
Storage Temperature: -2S°C to
+
70°C
(-13°F to
+
150°F)
Humidity: 0 to 90% RH (S°C to 30°C), non-
condensing
Cabinet: Supplied as model A (lightweight
metal cabinet) or C (as model A but with
flanges for standard 19
6
racks)
Dimensions: (A-cabinet without feet):
Height: 310,4mm (12,2in)
Width: 430mm (16,91n)
Depth: 500mm (19,7in)
Weight: 3S kg (771bs)
ACCESSORIES INCLUDED:
1 Test Unit .............................................. ZZ 0201
1 Mains cable ......................................... AN 0020
3 Signal cables for Test Unit and
2308 ......................................................... AO 0087
1 Power cable for Test Unit and
Remote control cable
for 2308 ................................................... AQ 0034
2 Spare fuses F2A/250V ..................... VF 0057
3 Spare fuses F 4A/250V ..................... VF OOS8
ACCESSORIES AVAILABLE:
Wide range of microphones and
accelerometers
IEC 62S-1 Interface cable (2m) ......... AO 0194
Adaptor for use with cable AO
0194 for connection Into an IEEE
488 Interface system ............................ AO 019S
3519 Sound Intensity Probe: For measure-
ment of Sound Intensity
2844 Line-Drive Amplifier: Miniature, charge
type amplifier for mounting directly on top of
accelerometers. Can be connected directly to
2032 Line Drive Input. 2032 Includes the nec-
essary line-drive supply
2313 Graphics Recorder: Includes BZ 7000
Application Package for direct copying of
display screen
BZ 7008 Application Package: For Type
2313. Extends the storage and measurement
capabilities of the system

Advertisement