300: Block_Logical_ Transfer 0, 16384 - outpl:JtS the DMA block from the main mem-
ory. For details of what the DMA block entails, refer to the 2032 main IEC
Interface Manual.
301: Block_Logical Transfer 17, 2048 - outputs the Time Function in Ch.A, or in Ch.B
Spectrum Averaging, the Time Function in Ch.B.
302: Block_Logical_ Transfer 18, 2048 - outputs the Time Function in Ch.B.
303: Block_Logical_ Transfer 19, 2048 - outputs the Autospectrum Ch.A, or in Ch.B
Spectrum Averaging, the Autospectrum in Ch.B.
304: Block_Logical_ Transfer 20, 2048 - outputs the Autospectrum Ch.B.
305: Block_Logical_ Transfer 21, 4096 - outputs the Cross Spectrum or, when Signal
·
Enhancement mode is in operation, the Enhanced Time Function Ch.B.
306: Block_Logical_ Transfer 19, 4096 - outputs the Enhanced Time Function Ch.A.
307: Block_Logical_ Transfer 23, 2048 - outputs the A-weighting function for the in-
put buffer, see the 2032 main IEC Interface Manual for details.
308: Block_Logical_ Transfer 30, 2048 - outputs the A-weighting function for the
stored buffer, see the 2032 main IEC Interface Manual for details.
1000*: Dump_Piot- means that the DIGITAL 1/0 "Output" pushkey has the same ac-
tion as the DIGITAL 1/0 "Plot" pushkey.
1001 *: causes output of the 2032 Main Cursor reading
1 002*: causes output of the 2032 Auxiliary Information field reading. Note, however,
that the Element Number and Flexible Cursor reading cannot be output.
1003*: Actuai_Function - outputs in ASCII the cursor values across the entire dis-
played trace containing the Field Selector.
1004 *: Actuai_Binary_Function - as Parameter 1003, except that the output is in bina-
ry.
30000: the selected text-string is output as an ASCII string, followed by the selected
interface terminator. This enables, for instance, programming instructions for
the 2032 to be written into the selected text-string and output to the 7400. A
series of instructions can then be assembled on the 7400 and re-entered into
the 2032 as a control program.
32767: selects the "learn" mode of the 2032 allowing output of a auto sequence to the
7400. The FIELD ENTRY knob must not be used when an autosequence is being
recorded. The procedure for recording an autosequence is as follows:
1. Press SYSTEM RESET 2 and set the Interface Terminator to 3. Set Special
Parameter 0 to 2032 and Special Parameter 80 to 32767.
2. Press SETUP LABEL "Display".
*
Data output using these modes cannot be re-entered into the 2032.